36V 10A Desktop Medical Power Adapter 2*Mopp Insulation Switching Power Supply IEC/UL 60601 UL CE Ukca Rcm

Still deciding? Get samples of $ !
Request Sample

Product Description

Basic Specifications

Model NO.APM360DB
CertificationTUV, RoHS, ISO9001, CCC, CE, Ukca, UL, Rcm, PSE, Kc
ColorBlack or White
TypeOEM
Input100~240VAC, 50/60Hz
Output Voltage12V~48V
Output Current7.0A~15.0A
Rate Power360W Max
Internal StructureOn-Off Type
ConnectionPlug In
Ripple & Noise100mv~350mv
SpecificationIEC/EN/UL 60601-1

Product Overview

This 360W (36V 10A) desktop medical power adapter is engineered for high-performance clinical and home healthcare systems requiring robust, safe, and reliable power. Fully compliant with IEC 60601-1:2012 (4th Edition) and global regional derivatives, it delivers stable, low-noise power while meeting the most stringent patient safety requirements. Featuring 2×MOPP (Means of Patient Protection) isolation, it provides dual independent insulation barriers, making it suitable for BF-type applied parts.

Key Features

  • 360W High-Power Output: Stable 36V / 10A continuous power for surgical robots, therapeutic lasers, and imaging peripherals.
  • 2×MOPP Isolation Certified: Exceeds basic safety requirements for direct or indirect patient contact.
  • Ultra-Low Leakage Current: <10 µA (normal), ensuring safety in demanding medical environments.
  • Reinforced Insulation: Tested at 4000 VAC for 1 minute for superior dielectric strength.
  • Global Compliance: UL, CE, UKCA, and RCM certifications under IEC 60601-1:2012.
  • Medical-Grade Protections: Includes SCP, OCP, OVP, OTP, and surge immunity.

Technical Specifications

Input ConnectorIEC 60320 C14
Output Accuracy36V DC ±1%
Isolation Rating2×MOPP
Applied PartBF Type
Efficiency≥90%
No-Load Power<0.5W
Operating Temp0°C to +50°C
Housing MaterialFlame-retardant Polycarbonate (UL94 V-0)

Model Selection Table

Model NO. Max Power Voltage Max Load Min Load Load Reg. Line Reg. Ripple
APM360DA300W24V12.5A0A±5%±1%100mV
APM360DB360W24V15.0A0A±5%±1%100mV
APM360DC300W36V8.33A0A±5%±1%100mV
APM360DD360W36V10.0A0A±5%±1%100mV
APM360DE300W48V6.25A0A±5%±1%100mV
APM360DF360W48V7.5A0A±5%±1%100mV
Medical Power Adapter Overview
Production Process
Quality Certification

Why 2×MOPP is Critical

Even within the Safety Extra-Low Voltage range, IEC 60601-1 requires 2×MOPP when circuits are patient-accessible. This ensures redundancy: if one insulation layer fails due to environmental stress, the second barrier prevents hazardous energy transfer. This adapter utilizes triple-insulated transformer windings and series Y-capacitors to maintain the highest safety margins.

Our Advantages

  • Automated Production: Fully automated lines operate 24/7, ensuring N times higher efficiency than manual assembly.
  • Consistent Quality: Robotic precision eliminates human error, resulting in ultra-low defect rates.
  • Flexible Delivery: Adjustment of capacity is fast, meeting peak-season demands with ease.
  • Global Standards: Full traceability and standardized production meeting European and American market requirements.

After-sales & Warranty

We provide a 24-month warranty for all AP series power adapters. Customers may request a return or exchange within 30 days of receipt for unused products in original packaging. Technical support, including installation guidance and troubleshooting, is available through our dedicated support channels.

Frequently Asked Questions

Q1: What does 2×MOPP isolation mean for this adapter? A: 2×MOPP (Means of Patient Protection) signifies dual independent insulation barriers between the hazardous mains voltage and the patient, complying with the strictest medical safety standards (IEC 60601-1).
Q2: Is this power supply suitable for home healthcare equipment? A: Yes, it is designed for both clinical and home healthcare systems, meeting medical-grade safety and EMI requirements.
Q3: What certifications are held by this product? A: It is fully certified with UL, CE, UKCA, RCM, TUV, RoHS, and more, specifically under the IEC 60601-1:2012 medical standard.
Q4: Can it handle sensitive medical imaging equipment? A: Yes, with low ripple and noise (≤200 mV), it is designed to ensure signal integrity in sensitive analog and digital circuits.
Q5: What is the patient leakage current? A: The adapter features ultra-low leakage current: <10 µA under normal conditions and <50 µA under single-fault conditions.
Q6: What is the standard warranty period? A: We offer a comprehensive 24-month warranty covering defects in materials and workmanship.

Related Products